Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Angiodynamics, Inc. recalls BioFlo DuraMax 15.5F 24cm Single Valve Sheath VascPak Kit- Indicated for use in attaining Long-Term vascular a

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.
Products
  • BioFlo DuraMax 15.5F 24cm Single Valve Sheath VascPak Kit- Indicated for use in attaining Long-Term vascular a
Where it was sold

VA

States: Virginia

Barcodes
25051684025680
Units affected
250 units
About this recall
BioFlo DuraMax 15.5F 24cm Single Valve Sheath VascPak Kit- Indicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis. Product Number: H965103038181