Argon Medical Devices, Inc recalls
Argon Medical Devices, Inc is named on 5 recalls in Recall Relay's records since November 2023, from FDA (11 recalled products: FDA lists each product as its own recall); the most recent was June 15, 2026.
All 11 recalled products, newest first.
-
Class IIofficial
Argon Medical Devices, Inc recalls L-Cath PICC S/L Peripherally Inserted Central Catheter: 26ga (1.9F) 0.60mm x 30cmCatheters, indicated for short/long-term peripheral access to central venous system for the administration of fluids, medications, nutrients; and sampling, may experience leaking and/or cracking at the hub, which may lead to replacement, inability to flush/aspirate catheter, catheter occlusion, therapy interruption, blood loss, tissue irritation, bloodstream infection.
-
Class IIofficial
Argon Medical Devices, Inc recalls L-Cath Midline S/L Catheter 20ga (3F) 1mm x 20cmCatheters, indicated for short/long-term peripheral access to central venous system for the administration of fluids, medications, nutrients; and sampling, may experience leaking and/or cracking at the hub, which may lead to replacement, inability to flush/aspirate catheter, catheter occlusion, therapy interruption, blood loss, tissue irritation, bloodstream infection.
-
Class Iofficial
Argon Medical Devices, Inc recalls Bone Marrow TrayKits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
-
Class Iofficial
Argon Medical Devices, Inc recalls T-Lok Bone Marrow TrayKits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
-
Class Iofficial
Argon Medical Devices, Inc recalls One-Piece Introducer TrayKits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
-
Class Iofficial
Argon Medical Devices, Inc recalls Transvenous Pacing Electrode Tray w/ Bipolar Balloon Pacing Electrode 5FKits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
-
Class Iofficial
Argon Medical Devices, Inc recalls Arterial Line Kit 20ga x 6Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
-
Class Iofficial
Argon Medical Devices, Inc recalls General Biospy TrayKits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
-
Class IIofficial
Argon Medical Devices, Inc recalls Option"ELITE Vena Cava Filter System REF 352506070E UDI-DI code: 00886333217151 The Option"ELITE Filter is indDue to complaints of increased resistance when advancing the dilator within the introducer sheath. This issue could potential result in procedural delays and/or risk of venous intimal injuries if excessive force is applied
-
Class IIofficial
Argon Medical Devices, Inc recalls BioPince Ultra Full Core Biopsy Instrument: REF: REF: 360-1080-01Holes have been identified in biopsy instrument packaging trays compromising the sterile barrier, which may lead to cross contamination, or the introduction of micro-organisms into the access site, which could lead to an infectious process, bacteremia, or sepsis.
-
Class IIofficial
Argon Medical Devices, Inc recalls L-CATH PICC S/L Peripherally Inserted Central CatheterPeripherally Inserted Central Catheter (PICC) is labeled 26G/1.9Fr x 30cm but package contains incorrect 26G/1.9Fr x 50cm PICC, which could lead to procedural delay.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions Argon Medical Devices, Inc or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.
Get recall alerts