Argon Medical Devices, Inc recalls L-Cath PICC S/L Peripherally Inserted Central Catheter: 26ga (1.9F) 0.60mm x 30cm
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Catheters, indicated for short/long-term peripheral access to central venous system for the administration of fluids, medications, nutrients; and sampling, may experience leaking and/or cracking at the hub, which may lead to replacement, inability to flush/aspirate catheter, catheter occlusion, therapy interruption, blood loss, tissue irritation, bloodstream infection.
- Products
-
- L-Cath PICC S/L Peripherally Inserted Central Catheter: 26ga (1.9F) 0.60mm x 30cm
- Company
- Argon Medical Devices, Inc
- Where it was sold
-
US: AR, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, MA, MD, MI, MN, MO, MS, NC, ND, NJ, NV, NY, OR, TN, TX, UT, VA, WA. OUS: MYS, CAN, SAU
States: Arkansas, California, Colorado, Connecticut, Florida, Georgia, Iowa, Idaho, Illinois, Indiana, Kansas, Massachusetts, Maryland, Michigan, Minnesota, Missouri, Mississippi, North Carolina, North Dakota, New Jersey, Nevada, New York, Oregon, Tennessee, Texas, Utah, Virginia, Washington
- Units affected
- 6928
- About this recall
- L-Cath PICC S/L Peripherally Inserted Central Catheter: 26ga (1.9F) 0.60mm x 30cm, REF: 384539; 18ga (3.5F) 1.15mm x 60cm, REF: 384464; Basic Kit 28GA (1.2F) x 25cm 1 Lumen), REF: 384516; 20ga (3F) 1.00mm x 60cm, REF: 384463