Augustine Temperature Management, LLC recalls
Augustine Temperature Management, LLC is named on 4 recalls in Recall Relay's records since November 2023, from FDA (13 recalled products: FDA lists each product as its own recall); the most recent was December 16, 2024.
All 13 recalled products, newest first.
-
Class IIofficial
Augustine Temperature Management, LLC recalls Hot Dog PATIENT WARMINGThere is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.
-
Class IIofficial
Augustine Temperature Management, LLC recalls Hot Dog PATIENT WARMINGThere is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.
-
Class IIofficial
Augustine Temperature Management, LLC recalls Hot Dog PATIENT WARMINGThere is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.
-
Class IIofficial
Augustine Temperature Management, LLC recalls Augustine Surgical Inc. HOT DOG PATIENT WARMINGInstructions for use update on HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 pertaining to Power-on self-test error code "EA POST".
-
Class IIofficial
Augustine Temperature Management, LLC recalls Hillrom Temperature Management ControllerInstructions for use update on HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 pertaining to Power-on self-test error code "EA POST".
-
Class IIofficial
Augustine Temperature Management, LLC recalls Augustine Surgical Inc. HOT DOG PATIENT WARMINGInstructions for use update on HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 pertaining to Power-on self-test error code "EA POST".
-
Class IIofficial
Augustine Temperature Management, LLC recalls Augustine Surgical Inc. HotDog Return Electrode accessory connector cablesUsers were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact .
-
Class IIofficial
Augustine Temperature Management, LLC recalls Augustine Surgical Inc. HotDog Return Electrode accessory connector cablesUsers were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact .
-
Class IIofficial
Augustine Temperature Management, LLC recalls Augustine Surgical Inc. HOT DOG PATIENT WARMINGHotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
-
Class IIofficial
Augustine Temperature Management, LLC recalls Augustine Surgical Inc. HOT DOG PATIENT WARMINGHotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
-
Class IIofficial
Augustine Temperature Management, LLC recalls Hillrom Temperature Management ControllerHotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
-
Class IIofficial
Augustine Temperature Management, LLC recalls Augustine Surgical Inc. HOT DOG PATIENT WARMINGHotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
-
Class IIofficial
Augustine Temperature Management, LLC recalls Augustine Surgical Inc. HOT DOG PATIENT WARMINGHotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions Augustine Temperature Management, LLC or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.
Get recall alerts