Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Augustine Temperature Management, LLC recalls Hot Dog PATIENT WARMING

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.
Products
  • Hot Dog PATIENT WARMING
Where it was sold

US Nationwide distribution in the states of TX, AL, NV; AU, NZ.

States: Alabama, Nevada, Texas

Barcodes
00855913001824
Units affected
1 unit
About this recall
Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 127 cm (50"), REF US502, to prevent or treat hypothermia