Cadwell Industries Inc recalls
Cadwell Industries Inc is named on 3 recalls in Recall Relay's records since November 2023, from FDA; the most recent was August 18, 2026.
All 3, newest first.
-
Class IIofficial
Cadwell Industries Inc recalls Brand Name: IOMAX Product Name: IOMAX Cortical Module Model/Catalog Number: 190296-200 UDI-DI code: 0084006710Stimulator device contains a defective printed circuit board assembly (PCBA) that may cause the device to unintentionally deliver electrical stimulation to the patient, or allow an unintended electrical current to reach the patient.
-
Class IIofficial
Cadwell Industries Inc recalls Cascade IOMAX Cortical ModuleElectroneurodiagnostic medical device may contain a defective printed circuit board which may cause damage to the other internal components resulting in unintended electrical stimulation.
-
Class IIofficial
Cadwell Industries Inc recalls Cascade IOMAX Cortical ModuleThere is the potential that an electroneurodiagnostic medical device insulator may be installed incorrectly which could result in arcing to patient connections resulting in potential unintended shock to the patient.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions Cadwell Industries Inc or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.
Get recall alerts