Cadwell Industries Inc recalls Cascade IOMAX Cortical Module
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- There is the potential that an electroneurodiagnostic medical device insulator may be installed incorrectly which could result in arcing to patient connections resulting in potential unintended shock to the patient.
- Products
-
- Cascade IOMAX Cortical Module
- Company
- Cadwell Industries Inc
- Where it was sold
-
Worldwide - US Nationwide distribution in the state of MD and the countries of Mexico, Canada, Malaysia, India, Taiwan, Saudi Arabia, European Union, Switzerland, South Africa, Iraq, United Kingdom.
States: Maryland
- Barcodes
00840067100653- Units affected
- 54 units
- About this recall
- Cascade IOMAX Cortical Module, REF: 190296-200