Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Cadwell Industries Inc recalls Cascade IOMAX Cortical Module

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
There is the potential that an electroneurodiagnostic medical device insulator may be installed incorrectly which could result in arcing to patient connections resulting in potential unintended shock to the patient.
Products
  • Cascade IOMAX Cortical Module
Where it was sold

Worldwide - US Nationwide distribution in the state of MD and the countries of Mexico, Canada, Malaysia, India, Taiwan, Saudi Arabia, European Union, Switzerland, South Africa, Iraq, United Kingdom.

States: Maryland

Barcodes
00840067100653
Units affected
54 units
About this recall
Cascade IOMAX Cortical Module, REF: 190296-200