Civco Medical Instruments Co. Inc. recalls
Civco Medical Instruments Co. Inc. is named on 2 recalls in Recall Relay's records since November 2023, from FDA (9 recalled products: FDA lists each product as its own recall); the most recent was March 2, 2026.
All 9 recalled products, newest first.
-
Class IIofficial
Civco Medical Instruments Co. Inc. recalls eTRAX Needle System Starter Kit 12G (for Aurora Trackers)There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
-
Class IIofficial
Civco Medical Instruments Co. Inc. recalls eTRAX Needle Sensor - 16G(for Aurora Trackers)There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
-
Class IIofficial
Civco Medical Instruments Co. Inc. recalls eTRAX Needle Sensor - 12G (for Aurora Trackers)There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
-
Class IIofficial
Civco Medical Instruments Co. Inc. recalls eTRAX Needle System Starter Kit 14G (for Aurora Trackers)There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
-
Class IIofficial
Civco Medical Instruments Co. Inc. recalls eTRAX Needle Sensor - 14G(for Aurora Trackers)There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
-
Class IIofficial
Civco Medical Instruments Co. Inc. recalls eTRAX Needle System Starter Kit 16G(for Aurora Trackers)There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
-
Class IIofficial
Civco Medical Instruments Co. Inc. recalls eTRAX Needle Sensor - 18G (for Aurora Trackers)There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
-
Class IIofficial
Civco Medical Instruments Co. Inc. recalls eTRAX Needle System Starter Kit 18G (for Aurora Trackers)There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
-
Class IIofficial
Civco Medical Instruments Co. Inc. recalls BX2 Needle GuideDuring the attachment of the needle guide to the covered probe, the tight tolerance and fit of the needle guide can cause a pinhole breach in the transducer cover when latching the guide in place.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions Civco Medical Instruments Co. Inc. or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.
Get recall alerts