Civco Medical Instruments Co. Inc. recalls BX2 Needle Guide
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- During the attachment of the needle guide to the covered probe, the tight tolerance and fit of the needle guide can cause a pinhole breach in the transducer cover when latching the guide in place.
- Products
-
- BX2 Needle Guide
- Where it was sold
-
OUS Distribution to countries of: Japan, Taiwan, France, Georgia, Germany and Denmark
- Barcodes
00841436120050- Units affected
- 506 boxes
- About this recall
- BX2 Needle Guide, Part Number 644-094, tool for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers