Civco Medical Instruments Co. Inc. recalls BX2 Needle Guide

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
During the attachment of the needle guide to the covered probe, the tight tolerance and fit of the needle guide can cause a pinhole breach in the transducer cover when latching the guide in place.
Products
  • BX2 Needle Guide
Where it was sold

OUS Distribution to countries of: Japan, Taiwan, France, Georgia, Germany and Denmark

Barcodes
00841436120050
Units affected
506 boxes
About this recall
BX2 Needle Guide, Part Number 644-094, tool for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers