DeRoyal Industries Inc recalls
DeRoyal Industries Inc is named on 7 recalls in Recall Relay's records since November 2023, from FDA (10 recalled products: FDA lists each product as its own recall); the most recent was August 20, 2026.
All 10 recalled products, newest first.
-
Class IIofficial
DeRoyal Industries Inc recalls REF: 81-030415 Myocardial Probe Non-PyrogenicMyocardial temperature probes contain an incorrect needle. Specifically, an 18mm needle was included instead of a 15mm needle.
-
Class IIofficial
DeRoyal Industries Inc recalls DeRoyal Industries Inc. Ref 46-406Sterile wound dressing, lacks sterility assurance
-
Class IIofficial
DeRoyal Industries Inc recalls DeRoyal REF 71-C5100NS TubingDue to the inner diameter of the suction tubing connector being smaller than the suction canister lid port connection. This size difference, combined with the durometer of the plastic material, makes the assembly to a suction canister lid port difficult. This could result in inadequate suction and cause a delay in patient care.
-
Class IIofficial
DeRoyal Industries Inc recalls DeRoyal AMG Suction Connector Tubing REF: 71-2083The tubing may not ensure that the yankauer stays intact, and secure to the tubing / yankauer connection for a tight seal. This could result in inadequate suction and cause delay in patient care.
-
Class IIofficial
DeRoyal Industries Inc recalls DeRoyal C-SECTION TRACEPACKDeRoyal is recalling certain lots of surgical procedure pack products that contain a Stryker SafeAir Smoke Evacuation Pencil. The recall is due to Stryker receiving reports that the pencil is activating without manual input when plugged into a power source or remains active after buttons are released.
-
Class IIofficial
DeRoyal Industries Inc recalls DeRoyal CHILDREN'S GENERAL TRACEPACKDeRoyal is recalling certain lots of surgical procedure pack products that contain a Stryker SafeAir Smoke Evacuation Pencil. The recall is due to Stryker receiving reports that the pencil is activating without manual input when plugged into a power source or remains active after buttons are released.
-
Class IIofficial
DeRoyal Industries Inc recalls DeRoyal CHILDREN'S ORTHOPEDIC TRACEPACKDeRoyal is recalling certain lots of surgical procedure pack products that contain a Stryker SafeAir Smoke Evacuation Pencil. The recall is due to Stryker receiving reports that the pencil is activating without manual input when plugged into a power source or remains active after buttons are released.
-
Class IIofficial
DeRoyal Industries Inc recalls DeRoyal CHILDREN'S LAPTRACEPACKDeRoyal is recalling certain lots of surgical procedure pack products that contain a Stryker SafeAir Smoke Evacuation Pencil. The recall is due to Stryker receiving reports that the pencil is activating without manual input when plugged into a power source or remains active after buttons are released.
-
Class Iofficial
DeRoyal Industries Inc recalls GeoMed Custom TracecartsDeRoyal manufactured and distributed Surgical Tracecarts that contain a Nurse Assist 16FR Silver Urine Meter Foley which were subsequently recalled.
-
Class IIofficial
DeRoyal Industries Inc recalls DeRoyal CARDIAC CATH SET-UP KIT PGYBKKit were manufactured and distributed with Edwards Life Sciences TruWave Disposable pressure transducers which were subsequently recalled by Edwards Life Sciences.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.