DeRoyal Industries Inc recalls DeRoyal REF 71-C5100NS Tubing

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Due to the inner diameter of the suction tubing connector being smaller than the suction canister lid port connection. This size difference, combined with the durometer of the plastic material, makes the assembly to a suction canister lid port difficult. This could result in inadequate suction and cause a delay in patient care.
Products
  • DeRoyal REF 71-C5100NS Tubing
Where it was sold

U.S.: IL, MN, NC, PA, and VA O.U.S.: N/A

States: Illinois, Minnesota, North Carolina, Pennsylvania, Virginia

Barcodes
007497565787380074975657885100749756578967
Units affected
714 cases
About this recall