Class IIofficial
Etac A/s recalls Brand Name: Molift Product Name: Molift Mover 300 Model/Catalog Numbers: (1) M1501A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.
-
Class IIofficial
Etac A/s recalls Brand Name: Molift Product Name: Molift 2-point sling bars (Accessory to mobile hoists) Model/Catalog Numbers:A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.
-
Class IIofficial
Etac A/s recalls Brand Name: Molift Product Name: Molift Mover 205 Model/Catalog Number: (1) M14000A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.
-
Class IIofficial
Etac A/s recalls Brand Name: Molift Product Name: Molift 4-point sling bars (Accessory to mobile hoists) Model/Catalog Number:A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.