Etac A/s recalls Brand Name: Molift Product Name: Molift Mover 205 Model/Catalog Number: (1) M14000

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.
Products
  • Brand Name: Molift Product Name: Molift Mover 205 Model/Catalog Number: (1) M14000
Company
Etac A/s
Where it was sold

Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Poland, Qatar Spain, Sweden, Switzerland, UK.

States: Pennsylvania

Barcodes
05707995032270
Units affected
95 units