Ethicon Endo-Surgery Inc recalls
Ethicon Endo-Surgery Inc is named on 4 recalls in Recall Relay's records since November 2023, from FDA (7 recalled products: FDA lists each product as its own recall); the most recent was September 11, 2025.
All 7 recalled products, newest first.
-
Class IIofficial
Ethicon Endo-Surgery Inc recalls ECHELON LINEAR Cutters Reload 80 mm BlueThe devices were shipped unsterilized.
-
Class Iofficial
Ethicon Endo-Surgery Inc recalls Endopath Echelon Vascular White Reload for Advanced Placement Tip (35mmFirm has received an increase in reports regarding inadvertent instrument lockout during surgical procedures involving the device. If lockout occurs, the device will momentarily activate but will not cut or staple tissue, and additional steps will be required to open it and remove it from tissue. If user is unable to remove the device from the tissue, potential harms to patient include surgical delay, bleeding/hemorrhage, life threatening hemorrhage/hemorrhagic shock, and conversion to open therapy. Two adverse events received, including one patient death.
-
Class IIofficial
Ethicon Endo Surgery, LLC recalls PDSII¿VIO¿30IN(75CM)¿USP6-0(M0.7)¿D/A¿C-1Identified curing issues with the silicone during the needles manufacturing process.
-
Class IIofficial
Ethicon Endo Surgery, LLC recalls ETHIBOND EXCEL GRN 36IN(90CM) USP2-0(M3) D/A MHIdentified curing issues with the silicone during the needles manufacturing process.
-
Class IIofficial
Ethicon Endo Surgery, LLC recalls PROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRMIdentified curing issues with the silicone during the needles manufacturing process.
-
Class IIofficial
Ethicon Endo Surgery, LLC recalls PDS¿PLUS¿VIO¿30IN(75CM)¿USP6-0(M0.7)¿D/A¿C-1¿MPIdentified curing issues with the silicone during the needles manufacturing process.
-
Class IIofficial
Ethicon Endo-Surgery Inc recalls ENDOPATH ECHELON 45mm RELOADS WITH GRIPPING SURFACE TECHNOLOGY (GST)There is the potential for incomplete staple line formation during the firing sequence, without most or all staples, which may lead to insufficient tissue apposition that could require surgical intervention. This may result in bleeding, prolonged surgery, or surgical delay.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.