Ethicon Endo-Surgery Inc recalls Endopath Echelon Vascular White Reload for Advanced Placement Tip (35mm
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Firm has received an increase in reports regarding inadvertent instrument lockout during surgical procedures involving the device. If lockout occurs, the device will momentarily activate but will not cut or staple tissue, and additional steps will be required to open it and remove it from tissue. If user is unable to remove the device from the tissue, potential harms to patient include surgical delay, bleeding/hemorrhage, life threatening hemorrhage/hemorrhagic shock, and conversion to open therapy. Two adverse events received, including one patient death.
- Products
-
- Endopath Echelon Vascular White Reload for Advanced Placement Tip (35mm
- Company
- Ethicon Endo-Surgery Inc
- Where it was sold
-
US distribution nationwide. OUS distribution to EMEA, LATAM, APAC, and Canada.
Nationwide
- Barcodes
10705036014591- Units affected
- 678,526
- About this recall
- Endopath Echelon Vascular White Reload for Advanced Placement Tip (35mm, 4 row); Product code VASECR35