Ferno-Washington Inc recalls
Ferno-Washington Inc is named on 3 recalls in Recall Relay's records since November 2023, from FDA; the most recent was June 19, 2026.
All 3, newest first.
-
Class IIofficial
Ferno-Washington Inc recalls Power F2 Powered Fastening SystemProduct may experience accelerated roller wear and potential deformation due to interference between the roller assembly and rail. In the worst-case condition, rail mechanism may become lodged during operation and movement of the stretcher along the rail may no longer be possible through either powered or manual operation. May cause delays in transport of patient to healthcare facility.
-
Class IIofficial
Ferno-Washington Inc recalls Power F2 Powered Fastening SystemFirmware issue may result in intermittent loss or degradation of powered functionality when paired with compatible stretcher. The issue may lead to the system not operating in powered mode, or exhibiting reduced powered performance. Issue may result in interruption or delay in patient transport.
-
Class IIofficial
Ferno-Washington Inc recalls FERNO POWER X2 POWERED AMBULANCE COTThe fastening post hardware could loosen allowing the fastening post to detach from the cot
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions Ferno-Washington Inc or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.