Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Ferno-Washington Inc recalls Power F2 Powered Fastening System

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Firmware issue may result in intermittent loss or degradation of powered functionality when paired with compatible stretcher. The issue may lead to the system not operating in powered mode, or exhibiting reduced powered performance. Issue may result in interruption or delay in patient transport.
Products
  • Power F2 Powered Fastening System
Where it was sold

Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, ID, IL, KY, LA, ME, MN, MO, MT, NC, NE, NJ, NM, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Netherlands and Poland.

States: Alabama, Arkansas, Arizona, California, Colorado, Florida, Idaho, Illinois, Kentucky, Louisiana, Maine, Minnesota, Missouri, Montana, North Carolina, Nebraska, New Jersey, New Mexico, New York, Ohio, Oklahoma, Pennsylvania, South Carolina, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin, West Virginia

Barcodes
0019079000657000190790003210
Units affected
319
About this recall
Power F2 Powered Fastening System, Power F2, Item No. 0567093 used with Power X2 Stretchers, Power X2, Item No. 0015816 and PowerX2 Premium, Item No. 0015821