Fresenius Medical Care Holdings, Inc. recalls
Fresenius Medical Care Holdings, Inc. is named on 19 recalls in Recall Relay's records since November 2023, from FDA (35 recalled products: FDA lists each product as its own recall); the most recent was May 12, 2026.
All 35 recalled products, newest first.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls Bicarby" Dialysate; Model Numbers: RFP-400-GGiven the increased number of leaks reported by users, it was determined to be the appropriate course of action to notify users of the potential for the occurrence of Bicarby luer-lock leaks during manipulation of the frangible and tubing when preparing the product for use for treatment. This issue presents the potential of a slip and fall hazard associated with fluid on the floor.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls 5008X Blood Tubing SetPotential for Online Port Errors due to connection issues when connecting the SafeLine of the 5008X CAREset to the substitution fluid port of the 5008X CAREsystem.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls Bicarby Dialysate; Model number: RFP-400-G;There is an error in the labeling of the GTIN on the product. When scanned; however, the case labeling, lot number, and printed labeling on the product correctly identifies the product code and the solution contained in the bag. In the event the incorrect GTIN labeling is used as the singular source for product identification, unrecognized use of a potassium solution with a concentration differing from the prescription may result.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls DELFLEXLack of Assurance of Sterility: Potential leaks from perforations in bags.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls DELFLEXLack of Assurance of Sterility: Potential leaks from perforations in bags.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls 5008X CAREsystem +CLiC +CDX;Several software anomalies with the potential to impact patient treatment, of which one includes the possibility to result in treatment stop if certain conditions are fulfilled.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls Optiflux¿ High Flux E-beam Dialyzer Model Numbers: (1) 0500316E OPTIFLUX F160NRe 12/CS 1.5SA EBEAMTo provide further information to the user to ensure proper use of the medical device. the caps had been changed from a threaded version to a press-on version whereby the user must press the caps firmly onto the dialyzer to securely affix the cap.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls 0.9% Sodium Chloride InjectionLack of Assurance of Sterility
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls 5008X HDF Hemodialysis Blood Tubing SetPotential of blood leak occurring between the lower port of the venous chamber and the tubing include: hemodiafiltration (HDF), hemodialysis (HD), hemofiltration (HF), and isolated ultrafiltration.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls Dialyzer Optiflux 160NRePotential for internal blood leaks due to cracked polyurethane
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls Blood Pressure Module BPM3 TM-2917B (Spare Part Numbers 191378 or 191389) with firmware version S005 installedMachines equipped with BPM3 Blood Pressure Modules (Blood pressure monitor TM-2917B with firmware version S005) will not record Blood Pressure measurements and does not alarm if the cuff tubing is kinked close to the cuff.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls DELFLEX PERITONEAL DIALYSIS SOLUTION With LOW MAGNESIUM / LOW CALCIUM 4.25% DEXTROSE and attached stay "safe EThis product is being recalled due to the tube weld failure presents itself as a slow leak and can be difficult to detect.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls RTL190828: 2008K@HOME HEMODIALYSIS SYSTEM - Blood Pump Rotor spare partAn increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls 191014: 2008K@HOME HD SYSTEM W/BIBAG (CANADA) - Blood Pump Rotor spare partAn increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls RTL190904: 2008K@HOME HEMODIALYSIS SYSTEM W/ BIBAG - Blood Pump Rotor spare partAn increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls 190828: 2008K@HOME HEMODIALYSIS SYSTEM - Blood Pump Rotor spare partAn increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls 190395: 2008K@HOME MACHINEAn increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls 190904 2008K@HOME HEMODIALYSIS SYSTEM W/ BIBAG - Blood Pump Rotor spare partAn increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls RTLR190395: 2008K@HOME MACHINEAn increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls RTL190395: 2008K@HOME MACHINEAn increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic PartConcentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is too high and stops the dialysis water supply.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls 2008T BlueStar Hemodialysis MachineAn increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls 2008T BlueStar Hemodialysis MachineAn increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls 2008T BlueStar Hemodialysis MachineAn increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls 2008T BlueStar Hemodialysis MachineAn increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls Blood Pump RotorAn increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls Novalung sensor box is a component of the Novalung Console (F30000162)-Indicated for long-term (> 6 hours) resIssue related to Novalung sensor box, which is a component of the Novalung Console (Part Number F30000163). In certain instances, error messages #206 (yellow) and #208 (red) technical failure, flow measurement during use of the Novalung system.
-
Class Iofficial
Fresenius Medical Care Holdings, Inc. recalls stay-safe /Safe Lock Catheter Ext. 12 inThe peroxide cross-linked tubing has leachables identified as NDL PCBAs.
-
Class Iofficial
Fresenius Medical Care Holdings, Inc. recalls stay-safe /Luer Lock Catheter Ext. 12 inThe peroxide cross-linked tubing has leachables identified as NDL PCBAs.
-
Class Iofficial
Fresenius Medical Care Holdings, Inc. recalls stay-safe /Luer Catheter Ext. 6 inThe peroxide cross-linked tubing has leachables identified as NDL PCBAs.
-
Class Iofficial
Fresenius Medical Care Holdings, Inc. recalls stay-safe /Luer Lock Catheter Ext. 18 inThe peroxide cross-linked tubing has leachables identified as NDL PCBAs.
-
Class Iofficial
Fresenius Medical Care Holdings, Inc. recalls stay-safe /Luer Lock Adapter 4 inThe peroxide cross-linked tubing has leachables identified as NDL PCBAs.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls DELFLEX Peritoneal Dialysis Solution in Biofine container 1.5% DextroseLack of Sterility Assurance
-
Class IIIofficial
Fresenius Medical Care Holdings, Inc. recalls Liberty Select CyclerAffected Liberty Select cyclers may display an incorrect time stamp on the treatment data report sent to the clinic.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls Stay Safe CapUpdating the labeling to include additional warnings and contraindications for potential side effects.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.