FDAMedical devicesRecalled Status: Ongoing

Fresenius Medical Care Holdings, Inc. recalls Blood Pressure Module BPM3 TM-2917B (Spare Part Numbers 191378 or 191389) with firmware version S005 installed

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Machines equipped with BPM3 Blood Pressure Modules (Blood pressure monitor TM-2917B with firmware version S005) will not record Blood Pressure measurements and does not alarm if the cuff tubing is kinked close to the cuff.
Products
  • Blood Pressure Module BPM3 TM-2917B (Spare Part Numbers 191378 or 191389) with firmware version S005 installed
Where it was sold

US Nationwide. Canada.

Nationwide

Units affected
3,718 systems
About this recall
Blood Pressure Module BPM3 TM-2917B (Spare Part Numbers 191378 or 191389) with firmware version S005 installed in 2008 Series Hemodialysis Machines with Part Numbers 190573, 190713, 190766, 190858, 190895, 190908, 191124, 191126, 191128, and 191130