In2bones USA, LLC recalls
In2bones USA, LLC is named on 2 recalls in Recall Relay's records since November 2023, from FDA; the most recent was September 9, 2026.
All 2, newest first.
-
Class IIofficial
In2bones USA, LLC recalls REF: P65 ST862 CoLag SCREWDue to labeling error, bone screws specified as fully threaded, are actually partially threaded screws.
-
Class IIofficial
In2bones USA, LLC recalls NeoSpan¿ SuperElastic Compression Staple w/Instruments Catalog Number and Product Name T50 SN010 NeoSpan¿ SupeDue to the Magnetic Resonance Conditional Symbol missing on the carton, inner tray, and patient label.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.