Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

In2bones USA, LLC recalls REF: P65 ST862 CoLag SCREW

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Due to labeling error, bone screws specified as fully threaded, are actually partially threaded screws.
Products
  • REF: P65 ST862 CoLag SCREW
Where it was sold

US Nationwide distribution in the states of FL, GA, HI, NC, OH, PA, SD, TX, VA.

States: Florida, Georgia, Hawaii, North Carolina, Ohio, Pennsylvania, South Dakota, Texas, Virginia

Barcodes
00817906023992
Units affected
29 units
About this recall
REF: P65 ST862 CoLag SCREW, FT, 6.7X62.5 mm