Inari Medical - Oak Canyon recalls
Inari Medical - Oak Canyon is named on 2 recalls in Recall Relay's records since November 2023, from FDA; the most recent was April 8, 2025.
All 2, newest first.
-
Class IIofficial
Inari Medical - Oak Canyon recalls Artix MT Thrombectomy DeviceThrombectomy device specifies vessel range of 2.5-6 mm on the pouch, carton, and Instructions for Use rather than the correct vessel range of 3-6 mm, which could incrementally increase the radial force on the artery, which has the potential of vascular complications.
-
Class Iofficial
Inari Medical - Oak Canyon recalls ClotTriever XL CatheterThrombectomy system catheter instructions for use updated to reduce adverse events due to the following circumstances: 1) Fibrous, firmly adherent, calcified material, or suspected tumor thrombus overloading catheter bag 2) Catheter pulled through internal jugular vein access site or not pulled away from heart; may lead to device entrapment, pulmonary embolism, venous cutdown, abdominal surgery.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions Inari Medical - Oak Canyon or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.
Get recall alerts