Inari Medical - Oak Canyon recalls Artix MT Thrombectomy Device
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Thrombectomy device specifies vessel range of 2.5-6 mm on the pouch, carton, and Instructions for Use rather than the correct vessel range of 3-6 mm, which could incrementally increase the radial force on the artery, which has the potential of vascular complications.
- Products
-
- Artix MT Thrombectomy Device
- Company
- Inari Medical - Oak Canyon
- Where it was sold
-
US Nationwide distribution in the states of FL, IL, WI, KY, TX, CA, OH, GA, AK, MI, PA, WA, CO, OH.
States: Alaska, California, Colorado, Florida, Georgia, Illinois, Kentucky, Michigan, Ohio, Pennsylvania, Texas, Washington, Wisconsin
- Barcodes
00810123710584- Units affected
- 45
- About this recall
- Artix MT Thrombectomy Device, REF: 32-102