Inpeco S.A. recalls
Inpeco S.A. is named on 3 recalls in Recall Relay's records since November 2023, from FDA (5 recalled products: FDA lists each product as its own recall); the most recent was October 30, 2025.
All 5 recalled products, newest first.
-
Class IIofficial
Inpeco S.A. recalls FlexLab (FLX); Version: FLX-217-10;The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results. This error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes.
-
Class IIofficial
Inpeco S.A. recalls FlexLab (FLX); Version: FLX-217-40;The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results, this error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes
-
Class IIofficial
Inpeco S.A. recalls FlexLab (FLX) System. Potassium Test System. in vitro diagnosticThe FlexLab System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the FDA, so its use may cause an increase in erroneous results or a delay in patient results.
-
Class IIofficial
Inpeco S.A. recalls FlexLab X (FXX) System. Potassium Test System. in vitro diagnosticThe FlexLab X System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the FDA, so its use may cause an increase in erroneous results or a delay in patient results.
-
Class IIofficial
Inpeco S.A. recalls Roche Cobas 8000 and Cobas PRO Interface Module (CO8 IM)- IVD FlexLab Automation System -a modular system desiFirmware of the Interface Module with Roche Cobas 8000 and Cobas PRO (CO8 IM) may lead to a delayed sample tube processing.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.