Inpeco S.A. recalls FlexLab (FLX); Version: FLX-217-40;
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results, this error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes
- Products
-
- FlexLab (FLX); Version: FLX-217-40;
- Company
- Inpeco S.A.
- Where it was sold
-
US Domestic: MI. OUS International: Australia, Finland, France, Germany, Italy, Portugal, Spain, and UK.
States: Michigan
- Barcodes
07640172344002- Units affected
- 14 units (1 US, 13 OUS)
- About this recall
- FlexLab (FLX); Version: FLX-217-40;