Integra LifeSciences Corp. recalls
Integra LifeSciences Corp. is named on 14 recalls in Recall Relay's records since November 2023, from FDA (57 recalled products: FDA lists each product as its own recall); the most recent was February 6, 2025.
All 57 recalled products, newest first.
-
Class IIofficial
Integra LifeSciences Corp. recalls AURORA Surgiscope SystemPossibility for the obturator to break (separate).
-
Class IIofficial
Integra LifeSciences Corp. recalls Brand Name: MediHoney¿ Gel Product Name: MediHoney¿ Wound & Burn Dressing Gel Model/Catalog Number: 31815 SoftPotential that the induction seal is not completely sealed to the device tube packaging.
-
Class IIofficial
Integra LifeSciences Corp. recalls CODMAN CERTAS Plus Programmable Valve Inline Small Valve Only. Model Number: 828810PL. Implantable device thatPotential for certain SKUs and lot numbers were distributed with the incorrect product description on the internal and external labels.
-
Class IIofficial
Integra LifeSciences Corp. recalls CODMAN CERTAS Plus Programmable Valve Right Angle Valve Only. Model Number: 828820. Implantable device that prPotential for certain SKUs and lot numbers were distributed with the incorrect product description on the internal and external labels.
-
Class IIofficial
Integra LifeSciences Corp. recalls CODMAN CERTAS Plus Programmable Valve Inline Small Valve Only. Model Number: 828810 Implantable device that prPotential for certain SKUs and lot numbers were distributed with the incorrect product description on the internal and external labels.
-
Class IIofficial
Integra LifeSciences Corp. recalls Duraform Dural Graft ImplantOut of specification endotoxin levels.
-
Class IIofficial
Integra LifeSciences Corp. recalls SURG PAT XRAY 1X1 Model/Catalog Number: 801403. The surgical patties and surgical strips are indicated for theIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls SURG STRP 1/4X6 Model/Catalog Number: 801450. The surgical patties and surgical strips are indicated for the uIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls SURG PAT XRAY 1/2X3 Model/Catalog Number: 801407. The surgical patties and surgical strips are indicated for tIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls SURG PAT XRAY 1/4X1-1/2 Model/Catalog Number: 801397. The surgical patties and surgical strips are indicated fIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls MICR PATIE RND Model/Catalog Number: 801396. CODMAN Surgical Patties and CODMAN Surgical Strips are manufacturIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls SURG STRP 1X6-200 Model/Catalog Number: 801453. The surgical patties and surgical strips are indicated for theIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls SURG PAT XRAY 1/2X1 Model/Catalog Number: 801402. The surgical patties and surgical strips are indicated for tIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls SURG STRP 3-1/2X6 Model/Catalog Number:801457. The surgical patties and surgical strips are indicated for theIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls SURG PAT XRAY 1/4X3 Model/Catalog Number: 801398. The surgical patties and surgical strips are indicated for tIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls SURG STRP 3X6 Model/Catalog Number: 801456. The surgical patties and surgical strips are indicated for the useIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls SURG STRP 2X6 Model/Catalog Number: 801455. The surgical patties and surgical strips are indicated for the useIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls SURG STRP 3/4X6 Model/Catalog Number: 801452. The surgical patties and surgical strips are indicated for the uIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls SURG PAT XRAY 1/2X1/2 Model/Catalog Number: 801400. The surgical patties and surgical strips are indicated forIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls SURG STRP 1/8X6 Model/Catalog Number: 801449. The surgical patties and surgical strips are indicated for the uIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls SURG STRP 1/2X6 Model/Catalog Number: 801451. The surgical patties and surgical strips are indicated for the uIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls SURG PAT XRAY 1/4X1/4 Model/Catalog Number: 801399. The surgical patties and surgical strips are indicated forIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls SURG PAT XRAY 3/4X3/4 Model/Catalog Number: 801401. The surgical patties and surgical strips are indicated forIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls SURG PAT XRAY 1/2X1-1/2 Model/Catalog Number: 801404. The surgical patties and surgical strips are indicated fIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls SURG PAT XRAY 1X3 Model/Catalog Number: 801408. The surgical patties and surgical strips are indicated for theIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls SURG PAT XRAY 3X3 Model/Catalog Number: 801409. The surgical patties and surgical strips are indicated for theIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls SURG STRP 1-1/2X6 Model/Catalog Number: 801454. The surgical patties and surgical strips are indicated for theIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls SURG PAT XRAY 1/2X2 Model/Catalog Number: 801406. The surgical patties and surgical strips are indicated for tIdentified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
-
Class IIofficial
Integra LifeSciences Corp. recalls MEDIHONEY Dressing with Active Leptospermum Honey Wound and Burn Dressing W/APPLICATORThere is a potential for pinholes in the applicator pouch film which creates a sterility concern.
-
Class IIofficial
Integra LifeSciences Corp. recalls TruDi NAV SuctionSpecific lots of TruDi NAV Suction Instruments 0 Degrees were incorrectly calibrated which when used with the TruDi Navigation System, may result in a discrepancy between the actual tip location and the visualized location displayed on the system.
-
Class IIofficial
Integra LifeSciences Corp. recalls AURORA Surgiscope System. SterilePotential sheath damage that could lead to tissue damage and/or track hemorrhage.
-
Class IIofficial
Integra LifeSciences Corp. recalls CUSA Excel C2600 23KHz Straight Handpiece-Indicated for use in surgical procedures where fragmentationPotential for the CUSA Excel 23KHz Straight Handpiece housing to crack, may result in delay of treatment
-
Class IIofficial
Integra LifeSciences Corp. recalls CODMAN EVD BACTISEAL 1.9mm I.D. Catheter Set- indicated for gaining access to the ventricles of the brain andDefect in the external sterile packaging (pouch) leading to a gap or/and weakness of the seal may compromise the product sterility.
-
Class IIofficial
Integra LifeSciences Corp. recalls CODMAN BACTISEAL Clear EVD Catheter 1.9mm I.D. Catheter Set- indicated for gaining access to the ventricles ofDefect in the external sterile packaging (pouch) leading to a gap or/and weakness of the seal may compromise the product sterility.
-
Class IIofficial
Integra LifeSciences Corp. recalls CODMAN BACTISEAL EVD Catheter Set - Indicated for gaining access to the ventricles of the brain and can be useDefect in the external sterile packaging (pouch) leading to a gap or/and weakness of the seal that may compromise the product sterility.
-
Class IIofficial
Integra LifeSciences Corp. recalls Integra¿ Cranial access kit (without Prep solutions) Bit and Guard- Used during a ventriculostomy procedure. CPotential holes and tears in the sterile packaging (header bag) of the Cranial Access Kit may compromise the sterility of the device.
-
Class IIofficial
Integra LifeSciences Corp. recalls Ref No: INSHITHRZN / Integra Cranial Access Kit. For access to the subarachnoid space or the lateral ventricleA defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.
-
Class IIofficial
Integra LifeSciences Corp. recalls Ref No: INSHITH / Integra¿ Cranial Access Kit. For access to the subarachnoid space or the lateral ventriclesA defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.
-
Class IIofficial
Integra LifeSciences Corp. recalls Ref No: HITHSP06 / Integra Cranial Access Kit. For access to the subarachnoid space or the lateral ventriclesA defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.
-
Class IIofficial
Integra LifeSciences Corp. recalls Ref No: INSHITHND / Integra¿ Cranial access kit (without Prep solutions) Bit and Guard. For access to the subaA defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.
-
Class IIofficial
Integra LifeSciences Corp. recalls INS7280 / Integra¿ Cranial Access Kit. For access to the subarachnoid space or the lateral ventricles of the bA defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.
-
Class IIofficial
Integra LifeSciences Corp. recalls Ref No: 31156 / COMBO KIT CONTAINS INS-8301A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.
-
Class IIofficial
Integra LifeSciences Corp. recalls Ref No: INS7250 / Integra Cranial Access Kit. For access to the subarachnoid space or the lateral ventricles oA defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.
-
Class IIofficial
Integra LifeSciences Corp. recalls Ref No: INS7270 / Integra¿ Cranial Access Kit. For access to the subarachnoid space or the lateral ventriclesA defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.
-
Class IIofficial
Integra LifeSciences Corp. recalls Ref No: INS7260 / Integra¿ Cranial Access Kit. For access to the subarachnoid space or the lateral ventriclesA defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.
-
Class IIofficial
Integra LifeSciences Corp. recalls Ref No: HITHSP04 / Integra Cranial Access Kit. For access to the subarachnoid space or the lateral ventriclesA defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.
-
Class IIofficial
Integra LifeSciences Corp. recalls Ref No: INS5HND / Integra Cranial access kit (without Prep solutions) Bit and Guard. For access to the subaracA defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.
-
Class IIofficial
Integra LifeSciences Corp. recalls Combo Kit Consists OF 10-110/INS-4500/INS-7040- For access to the subarachnoid space or the lateral ventriclesDefect in the outer packaging of the Cranial Access Kits. can cause the packaging to split without any additional forces outside compromising the sterility of the device
-
Class IIofficial
Integra LifeSciences Corp. recalls Integra¿ Cranial Access Kit-For access to the subarachnoid space or the lateral ventricles of the brain. CatalDefect in the outer packaging of the Cranial Access Kits. can cause the packaging to split without any additional forces outside compromising the sterility of the device
-
Class IIofficial
Integra LifeSciences Corp. recalls Integra Cranial Access Kit- For access to the subarachnoid space or the lateral ventricles of the brain. CatalDefect in the outer packaging of the Cranial Access Kits. can cause the packaging to split without any additional forces outside compromising the sterility of the device
-
Class IIofficial
Integra LifeSciences Corp. recalls Integra Cranial Access Kit-For access to the subarachnoid space or the lateral ventricles of the brain. CataloDefect in the outer packaging of the Cranial Access Kits. can cause the packaging to split without any additional forces outside compromising the sterility of the device
-
Class IIofficial
Integra LifeSciences Corp. recalls Integra Cranial Access Kit -For access to the subarachnoid space or the lateral ventricles of the brain. CatalDefect in the outer packaging of the Cranial Access Kits can cause the packaging to split without any additional forces outside compromising the sterility of the device
-
Class IIofficial
Integra LifeSciences Corp. recalls Integra¿ Cranial Access Kit-For access to the subarachnoid space or the lateral ventricles of the brain. CatalDefect in the outer packaging of the Cranial Access Kits. can cause the packaging to split without any additional forces outside compromising the sterility of the device
-
Class IIofficial
Integra LifeSciences Corp. recalls Integra Cranial Access Kit-For access to the subarachnoid space or the lateral ventricles of the brain. CataloDefect in the outer packaging of the Cranial Access Kits. can cause the packaging to split without any additional forces outside compromising the sterility of the device
-
Class IIofficial
Integra LifeSciences Corp. recalls Integra Cranial access kit (without Prep solutions) Bit and Guard-For access to the subarachnoid space or theDefect in the outer packaging of the Cranial Access Kits can cause the packaging to split without any additional forces outside compromising the sterility of the device
-
Class IIofficial
Integra LifeSciences Corp. recalls Integra Cranial Access Kit - For access to the subarachnoid space or the lateral ventricles of the brain. CataDefect in the outer packaging of the Cranial Access Kits. can cause the packaging to split without any additional forces outside compromising the sterility of the device
-
Class Iofficial
Integra LifeSciences Corp. recalls CereLink ICP Extension Cable (Extension Cable)Firm is expanding the existing recall RES 90457 for the Cerelink, initiated in June 2022, to include all CereLink ICP Extension Cables. The updated corrective action includes a design change to the extension cable to address the "out of range readings" issue.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.