Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Integra LifeSciences Corp. recalls TruDi NAV Suction

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Specific lots of TruDi NAV Suction Instruments 0 Degrees were incorrectly calibrated which when used with the TruDi Navigation System, may result in a discrepancy between the actual tip location and the visualized location displayed on the system.
Products
  • TruDi NAV Suction
Where it was sold

The devices are only sold in the US. The following states have received the product: AL, AZ CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, PA, SC, SD, TN, TX, VA, VT, WI.

States: Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Georgia, Iowa, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Michigan, Minnesota, Missouri, Mississippi, North Carolina, North Dakota, Nebraska, New Jersey, New Mexico, New York, Ohio, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Virginia, Vermont, Wisconsin

Barcodes
10705031245877
Units affected
265 units
About this recall
TruDi NAV Suction, 0 Degrees; Model/Catalog Number: TDNS000Z. TruDi" NAV Suction Instruments are intended for use with the TruDi" Navigation System during surgical procedures in ENT medicine and skull base surgery