Magellan Diagnostics, Inc. recalls
Magellan Diagnostics, Inc. is named on 4 recalls in Recall Relay's records since November 2023, from FDA (6 recalled products: FDA lists each product as its own recall); the most recent was September 4, 2026.
All 6 recalled products, newest first.
-
Class IIIofficial
Magellan Diagnostics, Inc. recalls LeadCare II Blood Lead Test Kit. Model/Catalog Number: 70-6762.In kit sub-lot 2530M-01 the date of expiration printed on the outer labeling may indicate the wrong expiration date of July 22, 2026 while the correct expiration date is July 13, 2026. If used the system will display an error message and prevent the operator from running the test. Resulting in delayed results.
-
Class IIofficial
Magellan Diagnostics, Inc. recalls LeadCare Ultra Blood Lead Test Kit; Catalog Number: 70-8098; The LeadCare Ultra kit is provided with the folloCustomers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.
-
Class IIofficial
Magellan Diagnostics, Inc. recalls (1)(2) LeadCare¿ II Blood Lead Test KitCustomers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.
-
Class IIofficial
Magellan Diagnostics, Inc. recalls LeadCare¿ Plus Blood Lead Test Kit; Catalog Number: 82-0004; The LeadCare¿ Plus kit is provided with the folloCustomers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.
-
Class IIofficial
Magellan Diagnostics, Inc. recalls The LeadCare¿ II Blood Lead Test System relies on electrochemistry and a unique Sensor to detect lead in wholeCalibration Buttons being used in the kitting process were incorrectly labeled as Lot 2323M for lot 2333M. The use of mislabeled calibration Button with Lot #2333 can lead to delayed or erroneous high or erroneous low results.
-
Class IIofficial
Magellan Diagnostics, Inc. recalls LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the quantitative measureDuring an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may have inadequate cap-to-tube seals. Manufacturing with these caps may result in cap-to-tube seals that allow leaking of HCl or may result in cap-to-tube seals that allow evaporation of treatment reagent. The consequence of this would be a change in the volume of treatment reagent and a change in the ratios of either added patient sample or control material to treatment reagent. The result of this might be a falsely elevated result (e.g., a patient false positive [PATFP] result or a Controls Out of Range-High [COOR-Hi] result).
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions Magellan Diagnostics, Inc. or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.