Magellan Diagnostics, Inc. recalls The LeadCare¿ II Blood Lead Test System relies on electrochemistry and a unique Sensor to detect lead in whole
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Calibration Buttons being used in the kitting process were incorrectly labeled as Lot 2323M for lot 2333M. The use of mislabeled calibration Button with Lot #2333 can lead to delayed or erroneous high or erroneous low results.
- Products
-
- The LeadCare¿ II Blood Lead Test System relies on electrochemistry and a unique Sensor to detect lead in whole
- Company
- Magellan Diagnostics, Inc.
- Where it was sold
-
US Nationwide distribution.
Nationwide
- Units affected
- 1,774 units
- About this recall
- The LeadCare¿ II Blood Lead Test System relies on electrochemistry and a unique Sensor to detect lead in whole blood. The LeadCare¿ II Test Kit is provided with the following materials: " Sensors (2 Containers of 24 sensors each; 48 tests) " Treatment Reagent Tubes (2 packages of 24 tubes each; 48 tests) " Heparinized Capillary Tubes/Plungers (50) " Transfer Droppers (50) " Calibration Button (1) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)