Merge Healthcare, Inc. recalls
Merge Healthcare, Inc. is named on 4 recalls in Recall Relay's records since November 2023, from FDA (9 recalled products: FDA lists each product as its own recall); the most recent was July 22, 2026.
All 9 recalled products, newest first.
-
Class IIofficial
Merge Healthcare, Inc. recalls Merge PACS and Merge OrthoPACSSoftware issue may cause measurements to be displayed inaccurately.
-
Class IIofficial
Merge Healthcare, Inc. recalls Merge OrthoPACSThumbnails from a previously opened patient study may be displayed in the Thumbnail Bar while a different patient study is currently open.
-
Class IIofficial
Merge Healthcare, Inc. recalls Merge PACSThumbnails from a previously opened patient study may be displayed in the Thumbnail Bar while a different patient study is currently open.
-
Class IIofficial
Merge Healthcare, Inc. recalls Merge HemoMerge Healthcare has been informed by Schiller AG (Baar, Switzerland) of an issue with the Schiller ARGUS PB-3000 Patient Data Module (PDM) where certain units trigger an error message on the Merge Hemo Client (Pressure Sensor Defective) during non-invasive blood pressure (NIBP) measurements, specifically when the initial inflation pressure is set in the high-pressure range (>240 mmHg). The ARGUS PB-3000 is a vital signs monitoring instrument intended to be integrated into the Merge Hemo System. No other components of the Merge Hemo System are affected.
-
Class IIofficial
Merge Healthcare, Inc. recalls VERICISWhen following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
-
Class IIofficial
Merge Healthcare, Inc. recalls VERICISWhen following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
-
Class IIofficial
Merge Healthcare, Inc. recalls Merge CardioWhen following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
-
Class IIofficial
Merge Healthcare, Inc. recalls VERICISWhen following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
-
Class IIofficial
Merge Healthcare, Inc. recalls VERICISWhen following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions Merge Healthcare, Inc. or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.
Get recall alerts