Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Merge Healthcare, Inc. recalls VERICIS

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
Products
  • VERICIS
Where it was sold

US

Nationwide

Barcodes
00842000100768
Units affected
37
About this recall
VERICIS, Merge Cardio, Model/Catalog Number: Version 12.X, Software Version: 12.0.1, 12.0.2; The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.