Noxbox Ltd is named on 5 recalls in Recall Relay's records since November 2023, from FDA; the most recent was September 14, 2026.
All 5, newest first.
-
Class IIofficial
Noxbox Ltd recalls NOxBOXi Nitric Oxide Delivery System. Model Number: NOXBOX-I.Potential for irregular oxygen readings due to a manufacturing issue affecting certain oxygen sensors.
-
Class IIofficial
Noxbox Ltd recalls NOxBOXi Nitric Oxide Delivery System.Potential for oxygen tubing leaks due to failure of the tubing or tubing connection, which may lead to oxygen desaturation.
-
Class Iofficial
Noxbox Ltd recalls NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-IThe unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts.
-
Class Iofficial
Noxbox Ltd recalls NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-IFluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).
-
Class IIofficial
Noxbox Ltd recalls NOxBOXi Nitric Oxide Delivery SystemReplacement of the internal Sample Pump due to high failure rates.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.