Noxbox Ltd recalls NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).
- Products
-
- NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I
- Company
- Noxbox Ltd
- Where it was sold
-
Domestic: TN
States: Tennessee
- Barcodes
05060541640009- Units affected
- 1667 units
- About this recall
- NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I