Paragon 28, Inc. recalls
Paragon 28, Inc. is named on 4 recalls in Recall Relay's records since November 2023, from FDA (7 recalled products: FDA lists each product as its own recall); the most recent was July 10, 2026.
All 7 recalled products, newest first.
-
Class IIIofficial
Paragon 28, Inc. recalls Custom Made Device Implant 6123 Navicular Cage Nominal Right Rev ADevice was recalled due to an elevated oxygen level in excess of the allowable limit. Elevated oxygen levels may contribute to reduced material ductility.
-
Class IIIofficial
Paragon 28, Inc. recalls Custom Made Device Implant 6123 Navicular Cage Small Right Rev ADevice was recalled due to an elevated oxygen level in excess of the allowable limit. Elevated oxygen levels may contribute to reduced material ductility.
-
Class IIofficial
Paragon 28, Inc. recalls P28 PHANTON INTRAMEDULLARY NAIL SYSTEMParagon 28 received two (2) complaints were received involving breaches in the sterile barrier of the pouch packaging.
-
Class IIofficial
Paragon 28, Inc. recalls Phantom Fibula NailFibula nail system intended for use in fixation of fibular fractures and osteotomies, has out of specification thread depths on the fibula nail, so implant may not be able to mate properly with the associated mounting bolt. If nail (implant) cannot properly mate with instrumentation, the surgery potentially will need to be aborted.
-
Class IIofficial
Paragon 28, Inc. recalls Phantom TTC NailMedical devices distributed, prior to sterilization
-
Class IIofficial
Paragon 28, Inc. recalls Phantom TTC NailMedical devices distributed, prior to sterilization
-
Class IIofficial
Paragon 28, Inc. recalls Phantom TTC NailMedical devices distributed, prior to sterilization
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions Paragon 28, Inc. or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.
Get recall alerts