Paragon 28, Inc. recalls Phantom Fibula Nail
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Fibula nail system intended for use in fixation of fibular fractures and osteotomies, has out of specification thread depths on the fibula nail, so implant may not be able to mate properly with the associated mounting bolt. If nail (implant) cannot properly mate with instrumentation, the surgery potentially will need to be aborted.
- Products
-
- Phantom Fibula Nail
- Company
- Paragon 28, Inc.
- Where it was sold
-
US Nationwide distribution in the states of NJ, WA, NY, MD, CA, FL, NM, TX, IL, OH, NV, TN, LA.
States: California, Florida, Illinois, Louisiana, Maryland, New Jersey, New Mexico, Nevada, New York, Ohio, Tennessee, Texas, Washington
- Barcodes
00889795133670- Units affected
- 20
- About this recall
- Phantom Fibula Nail, 2.8mm x 130mm, Right, Ti, Fibula Nail, REF: P36-128-130R-S