Philips Medical Systems is named on 6 recalls in Recall Relay's records since November 2023, from FDA (12 recalled products: FDA lists each product as its own recall); the most recent was August 21, 2026.
All 12 recalled products, newest first.
-
Class IIofficial
Philips Medical Systems recalls Spectral CT and Spectral CT 7500 RT; Software Version Numbers: 5.2.X (U.S.); 5.0.XSoftware issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or unintended system motion.
-
Class IIofficial
Philips Medical Systems recalls Ingenuity CT CT5000 Ingenuity Plus CT5000 Ingenuity Pro CT5000 Ingenuity Premium Ingenuity CT Upgrades IngenuiThe clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.
-
Class IIofficial
Philips Medical Systems recalls Brilliance iCT Brilliance iCT Upgrades Brilliance iCT SP Brilliance CT 64 Channel CT6000 Brilliance 64 UpgradeThe clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.
-
Class IIofficial
Philips Medical Systems recalls IQon Spectral CT IQon Spectral CT UpgradesThe clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.
-
Class IIofficial
Philips Medical Systems recalls Philips Spectral CT on Rails. Model Number: 728334.Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of Previous Surview Exists Select Previous Surview? or should display but does not appear. 3. After performing Surview and planning the Brain Helical acquisition by setting Brain Area DoseRight Index to increase the dose, there is the potential that the Define Head Area option in the context menu is grayed out.
-
Class IIofficial
Philips Medical Systems recalls Model: Incisive CT; Product Code (REF): (1) 728143Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation), the WED (Water Equivalent Diameter) value might set itself to zero, and this may lead to an insufficient dose setting after the surview. If operator misses the insufficient dose and the WED value in User Interface and continues with the subsequent helical/Axial scans, then the obtained images will be noisy due to low dose setting. Issue 3: Due to a software failure, the ECG wave file is not saved. As a result, the cardiac offline image reconstruction fails due to the missing ECG wave file, and the user might rescan the patient. Issue 4: This issue is software fault in which the patient orientation will be changed to NULL under specific random scenarios resulting in radiation being delivered from the wrong orientation (surview scan) or in the wrong position (clinical scan) and will generate incorrect and undiagnosable surview/clinical images. Patients may be rescanned for surview and/or clinical images. Issue 5: During scanning, the preview image display is not consistent, the images are not arriving at a fixed rate, skipping images from time to time during scanning. The obtained images might not support making the clinical diagnosis and the user might decide to re-scan the patient. Issue 6: If a user presses the left and middle/right mouse buttons together or afterwards in a short time, both commands are executed and an incorrect auto ROI is created. This ROI coordinates won t be assigned as the object is empty. As a result, the threshold of contrast level will not be reached in the UI and this will prevent the system from automatically triggering the subsequent clinical scan. There is potential safety risk identified for additionally tracker shots and/or rescan of a patient when the clinical scan is not triggered. This issue only affects the manual ROI process.
-
Class IIofficial
Philips Medical Systems recalls Model: CT 5300; Product Code (REF): 728285; Software Versions: 5.1.0.X & 5.1.1.X;Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation), the WED (Water Equivalent Diameter) value might set itself to zero, and this may lead to an insufficient dose setting after the surview. If operator misses the insufficient dose and the WED value in User Interface and continues with the subsequent helical/Axial scans, then the obtained images will be noisy due to low dose setting. Issue 3: Due to a software failure, the ECG wave file is not saved. As a result, the cardiac offline image reconstruction fails due to the missing ECG wave file, and the user might rescan the patient. Issue 4: This issue is software fault in which the patient orientation will be changed to NULL under specific random scenarios resulting in radiation being delivered from the wrong orientation (surview scan) or in the wrong position (clinical scan) and will generate incorrect and undiagnosable surview/clinical images. Patients may be rescanned for surview and/or clinical images. Issue 5: During scanning, the preview image display is not consistent, the images are not arriving at a fixed rate, skipping images from time to time during scanning. The obtained images might not support making the clinical diagnosis and the user might decide to re-scan the patient. Issue 6: If a user presses the left and middle/right mouse buttons together or afterwards in a short time, both commands are executed and an incorrect auto ROI is created. This ROI coordinates won t be assigned as the object is empty. As a result, the threshold of contrast level will not be reached in the UI and this will prevent the system from automatically triggering the subsequent clinical scan. There is potential safety risk identified for additionally tracker shots and/or rescan of a patient when the clinical scan is not triggered. This issue only affects the manual ROI process.
-
Class IIofficial
Philips Medical Systems recalls Model: Incisive CT for Brazil SKD; Product Code (REF): 728146; Software Versions: 5.1.0.X & 5.1.1.X;Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axial scan with ORI (Dose Right index)/ DOM (Dose Modulation), the WED (Water Equivalent Diameter) value might set itself to zero, and this may lead to an insufficient dose setting after the surview. If operator misses the insufficient dose and the WED value in User Interface and continues with the subsequent helical/Axial scans, then the obtained images will be noisy due to low dose setting. Issue 3: Due to a software failure, the ECG wave file is not saved. As a result, the cardiac offline image reconstruction fails due to the missing ECG wave file, and the user might rescan the patient. Issue 4: This issue is software fault in which the patient orientation will be changed to NULL under specific random scenarios resulting in radiation being delivered from the wrong orientation (surview scan) or in the wrong position (clinical scan) and will generate incorrect and undiagnosable surview/clinical images. Patients may be rescanned for surview and/or clinical images. Issue 5: During scanning, the preview image display is not consistent, the images are not arriving at a fixed rate, skipping images from time to time during scanning. The obtained images might not support making the clinical diagnosis and the user might decide to re-scan the patient. Issue 6: If a user presses the left and middle/right mouse buttons together or afterwards in a short time, both commands are executed and an incorrect auto ROI is created. This ROI coordinates won t be assigned as the object is empty. As a result, the threshold of contrast level will not be reached in the UI and this will prevent the system from automatically triggering the subsequent clinical scan. There is potential safety risk identified for additionally tracker shots and/or rescan of a patient when the clinical scan is not triggered. This issue only affects the manual ROI process.
-
Class IIofficial
Philips Medical Systems recalls Spectral CT on Rails: Software Version 5.1.0.XA software issue that can incorrectly label all captured images that may lead to misdiagnosis and subsequent incorrect treatment.
-
Class IIofficial
Philips Medical Systems recalls Spectral CT 7500: Software Version 5.0A software issue that can incorrectly label all captured images that may lead to misdiagnosis and subsequent incorrect treatment.
-
Class IIofficial
Philips Medical Systems recalls CombiDiagnost R90 1.0 (709030)While performing a fluoroscopy examination, there is a potential that the Radio Fluoroscopy (RF) viewer will also display a previous patient's radiography images. If the issue occurs, there will be differences in image content, image format and image size.
-
Class IIofficial
Philips Medical Systems recalls CombiDiagnost R90 1.1 (709031)While performing a fluoroscopy examination, there is a potential that the Radio Fluoroscopy (RF) viewer will also display a previous patient's radiography images. If the issue occurs, there will be differences in image content, image format and image size.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.