Philips Medical Systems recalls Philips Spectral CT on Rails. Model Number: 728334.

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of Previous Surview Exists Select Previous Surview? or should display but does not appear. 3. After performing Surview and planning the Brain Helical acquisition by setting Brain Area DoseRight Index to increase the dose, there is the potential that the Define Head Area option in the context menu is grayed out.
Products
  • Philips Spectral CT on Rails. Model Number: 728334.
Where it was sold

Worldwide - US Nationwide distribution in the states of Minnesota and the countries of France and Netherlands.

States: Minnesota

Barcodes
00884838103627
Units affected
3 systems
About this recall
Philips Spectral CT on Rails. Model Number: 728334.