Raysearch Laboratories AB recalls
Raysearch Laboratories AB is named on 6 recalls in Recall Relay's records since November 2023, from FDA (44 recalled products: FDA lists each product as its own recall); the most recent was May 20, 2026.
All 44 recalled products, newest first.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayCare 2024A SP4. Oncology Information Systems.Beam set delivery note issues in which there is potential data loss if the beam set delivery note is modified simultaneously and/or if a note is updated during a treatment session, the user is not notified of the change.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayCare 2024A SP1. Oncology Information Systems.Beam set delivery note issues in which there is potential data loss if the beam set delivery note is modified simultaneously and/or if a note is updated during a treatment session, the user is not notified of the change.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation/RayPlan. Model/Catalog Number: 10.1. Software Version: RayStation/RayPlan 10B. Radiation Therapy TrPotential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
-
Class IIofficial
Raysearch Laboratories Ab recalls Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 10.1.1 Software Version: RayStatPotential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation/RayPlan. Model/Catalog Numbers: 11.0.0Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
-
Class IIofficial
Raysearch Laboratories Ab recalls Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 13.0.0.1547Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation/RayPlan. Model/Catalog Numbers: 12.0.0Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
-
Class IIofficial
Raysearch Laboratories Ab recalls Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 16.0.0 Software Version: RayStatPotential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
-
Class IIofficial
Raysearch Laboratories Ab recalls Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 15.2.0Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
-
Class IIofficial
Raysearch Laboratories Ab recalls Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 14.0.0Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
-
Class IIofficial
Raysearch Laboratories Ab recalls Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation with the following product descriptions: 1. RayStation 2023BDICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation with the following product descriptions: 1. RayStation 7DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation with the following product descriptions: 1. RayStation 12ADICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation with the following product descriptions: 1. RayStation 10BDICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation with the following product descriptions: 1. RayStation 6 Service Pack 1DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation with the following product descriptions: 1. RayStation 9BDICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation with the following product descriptions: 1. RayStation 11ADICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation with the following product descriptions: 1. RayStation 8BDICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation with the following product descriptions: 1. RayStation 11BDICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
-
Class IIofficial
Raysearch Laboratories Ab recalls Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 6.0.0.Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
-
Class IIofficial
Raysearch Laboratories Ab recalls Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 5.0.2.Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
-
Class IIofficial
Raysearch Laboratories Ab recalls Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 13.0.0Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
-
Class IIofficial
Raysearch Laboratories Ab recalls Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 6.1.1.Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
-
Class IIofficial
Raysearch Laboratories Ab recalls Brand Name: RayStation -Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 9.1.0.Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
-
Class IIofficial
Raysearch Laboratories Ab recalls Brand Name: RayStation Intended as a Radiation Therapy Treatment Planning System. Model/Catalog Number: 7.0.0.Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
-
Class IIofficial
Raysearch Laboratories Ab recalls Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 4.5.1.Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
-
Class IIofficial
Raysearch Laboratories Ab recalls Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 10.1.0Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
-
Class IIofficial
Raysearch Laboratories Ab recalls Brand Name: RayStation -Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 8.1.0.Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
-
Class IIofficial
Raysearch Laboratories Ab recalls Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 12.0.0Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
-
Class IIofficial
Raysearch Laboratories Ab recalls Brand Name: RayStation- Intended as a Radiation Therapy Treatment Planning System Model/Catalog Number: 1.0.0.Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation 9.1.0.933 and 9.2.0.483. Radiation Therapy Treatment Planning System.Potential for reported SSD to be too high.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation 14.0.0.3338. Radiation Therapy Treatment Planning System.Potential for reported SSD to be too high.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation 12.0.0.932Potential for reported SSD to be too high.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation 10.1.0.613 and 10.1.1.54. Radiation Therapy Treatment Planning System.Potential for reported SSD to be too high.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation 10.0.0.1154Potential for reported SSD to be too high.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation 7.0.0.19. Radiation Therapy Treatment Planning System.Potential for reported SSD to be too high.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation 15.0.0.430. Radiation Therapy Treatment Planning System.Potential for reported SSD to be too high.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation 8.0.0.61Potential for reported SSD to be too high.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation 11.0.0.951Potential for reported SSD to be too high.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation 9.0.0.113 and 9.0.1.142. Radiation Therapy Treatment Planning System.Potential for reported SSD to be too high.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation 8.1.0.47Potential for reported SSD to be too high.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation 13.0.0.1547Potential for reported SSD to be too high.
-
Class IIofficial
Raysearch Laboratories Ab recalls RayStation Software Version Numbers version numbers 8BA use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.