Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Raysearch Laboratories Ab recalls RayStation/RayPlan. Model/Catalog Numbers: 12.0.0

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
Products
  • RayStation/RayPlan. Model/Catalog Numbers: 12.0.0
Where it was sold

Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS

Barcodes
07350002010426
Units affected
22 units
About this recall
RayStation/RayPlan. Model/Catalog Numbers: 12.0.0, 12.1.0, 12.0.3, 12.0.4, 12.3.0. Software Version: RayStation 11B, RayStation 11B SP1, RayStation 11B SP2, RayStation 11B SP3, RayStation 11B SPT1. Radiation Therapy Treatment Planning System