Reflexion Medical, Inc. recalls
Reflexion Medical, Inc. is named on 3 recalls in Recall Relay's records since November 2023, from FDA; the most recent was February 9, 2026.
All 3, newest first.
-
Class IIofficial
Reflexion Medical, Inc. recalls RefleXion X1 Radiotherapy System Model RXM1000Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose.
-
Class IIofficial
Reflexion Medical, Inc. recalls RefleXion X1 Model RXM1000 UDI Codes: (01)00860003983812(11)230504(21)X11001 (01)00860003983812(11)230518(21)XDue to incorrect software configuration that potentially allows more than two (2) fractions within a 12-hour period (fractionLimits) and could potentially lead to adverse events (toxicity).
-
Class IIofficial
Reflexion Medical, Inc. recalls RefleXion X1 Radiotherapy System Model RXM1000 Software version 2.1.19-3 Radiation therapy system is a deviceA potential dose error exists for patients treated with an out of session SCINTIX partial fraction.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions Reflexion Medical, Inc. or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.
Get recall alerts