Reflexion Medical, Inc. recalls RefleXion X1 Radiotherapy System Model RXM1000
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose.
- Products
-
- RefleXion X1 Radiotherapy System Model RXM1000
- Company
- Reflexion Medical, Inc.
- Where it was sold
-
US: CA, TX, PA, CT, NJ, OR, LA, OH
States: California, Connecticut, Louisiana, New Jersey, Ohio, Oregon, Pennsylvania, Texas
- Barcodes
00860003983812- Units affected
- 12
- About this recall
- RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00