Reflexion Medical, Inc. recalls RefleXion X1 Radiotherapy System Model RXM1000

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose.
Products
  • RefleXion X1 Radiotherapy System Model RXM1000
Where it was sold

US: CA, TX, PA, CT, NJ, OR, LA, OH

States: California, Connecticut, Louisiana, New Jersey, Ohio, Oregon, Pennsylvania, Texas

Barcodes
00860003983812
Units affected
12
About this recall
RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00