Sedecal Sa recalls
Sedecal Sa is named on 4 recalls in Recall Relay's records since November 2023, from FDA (49 recalled products: FDA lists each product as its own recall); the most recent was June 10, 2025.
All 49 recalled products, newest first.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-40HF-B-D-C; 40KW CANON READYSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model SM-32HF-Batt; 32KW ANALOGSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-40HF-B-D-C; 40KW70CDEMO.005Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number 40KWFXSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-40HF-B-D-C; 40KW 50GSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-40HF-B-D-C; 40KW 55C/60CSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-40HF-Batt; 40KW ANALOGSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number 40KWFXPLUSSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-40HF-B-D-C; 40KW CANON READY.005Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-40HF-B-D-C; 40KW 70C ENSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-40HF-B-D-C; 40KW 50CSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls SED URS LPSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-40HF-B-D-C; 40KW 50GSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-40HF-B-D-C; 40KW 70C READYSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number 40KWFXPLUS-710CWSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-40HF-B-D-C; 40KW 55CSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model SM-40HF-B-D-C; 40KW 70CSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-40HF-B-D-C; 40KW 55CSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Sedecal Easy Moving Plus Model Number SM-50HF-B-D-C; 50KW 50GSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-40HF-B-D-C; 40KW 55GSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number 40KWFXPLUS.002Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-40HF-B-D-C; 40KW 80CSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number 40KWFXPLUS.005Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number 40KWMOBILE3.899Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Sedecal Mobile Digital Diagnostic X-Ray Systems; SM-40HF-B-D-KM (AeroDR X30)Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls RADPRO MOBILESedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls SM_40HF_BDCSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-40HF-B-D-C; 40KW 55G/60GSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-40HF-B-D-C; 40KW 55CSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number 40KWFXPLUS.889Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls SM_40HF_B_D_VIR3.004Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-40HF-B-D-C; 40KW.005Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model SM-40HF-B-D-C; 40KW - CANON READYSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number 40KWFXPLUS-710CW.007Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number 40KW-FX.899Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-40HF-B-D-C; 40KW.002Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls SHIMADZU RF.004Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-20HF-Batt; 20KW ANALOGSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls SM_40HF_BDC3Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls SM_40HF_B_D_VIR3Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-40HF-B-D-C; 40KW 70CSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-40HF-B-D-C; 40KWSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-40HF-B-D-C; 40KW 55C/60CSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number 40KWFXPLUS.004Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Model Number SM-32HF-B-D-C; 32KW 50GSedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls MobileDiagnost wDR 2.2; Model Number REF 9890-010-89522Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
-
Class IIofficial
Sedecal Sa recalls Battery Mobile X-ray Unit EASY MOVING Model with telescopic column 3.1Potential interference with the anti-fall system.
-
Class IIofficial
Sedecal Sa recalls Sedecal X Optima URS digital mobile diagnostic x-ray system. A a stationary x-ray unit with a universal swivelSome parts have fallen off the equipment resulting in a safety risk to patients and/or users.
-
Class IIofficial
Sedecal Sa recalls MobileDiagnost wDRWhile cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause a short circuit of the fine positioning control button. This would result in the system moving on its own in a very slow speed forward circular motion.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.