FDAMedical devicesRecalled Status: Ongoing

Sedecal Sa recalls MobileDiagnost wDR

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
While cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause a short circuit of the fine positioning control button. This would result in the system moving on its own in a very slow speed forward circular motion.
Products
  • MobileDiagnost wDR
Company
Sedecal Sa
Where it was sold

US Nationwide distribution. There was government distribution but no military distribution.

Nationwide

Barcodes
00884838028586008848380823730843604600217300884838091504
Units affected
851 devices
About this recall
MobileDiagnost wDR, Model numbers 712001, 712002, 712004, 712006, and 712007