Spacelabs Healthcare, Inc. recalls
Spacelabs Healthcare, Inc. is named on 8 recalls in Recall Relay's records since November 2023, from FDA (11 recalled products: FDA lists each product as its own recall); the most recent was June 4, 2026.
All 11 recalled products, newest first.
-
Class IIofficial
Spacelabs Healthcare, Inc. recalls Brand Name: Xhibit Product Name: Xhibit Central Station Model/Catalog Number: 96102 Software Version: N/A ProdDue to off-the-shelf display port component that has audio/video failures
-
Class IIofficial
Spacelabs Healthcare, Inc. recalls Brand Name: Spacelabs Ultraview SL Command Module Product Name: SL Command Module Model/Catalog Number: 91496Due to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester).
-
Class IIofficial
Spacelabs Healthcare, Inc. recalls Multi-parameter Command ModuleMulti-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.
-
Class IIofficial
Spacelabs Healthcare, Ltd. recalls Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 ProduDue to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record
-
Class IIofficial
Spacelabs Healthcare, Inc. recalls Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 01Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.
-
Class IIofficial
Spacelabs Healthcare, Inc. recalls Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 01Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.
-
Class IIofficial
Spacelabs Healthcare, Inc. recalls Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 01Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.
-
Class IIofficial
Spacelabs Healthcare, Inc. recalls Xhibit Telemetry Receiver with SoftwareTelemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts and necessary clinical intervention could be delayed.
-
Class IIofficial
Spacelabs Healthcare, Ltd. recalls Eclipse Mini REF 98900 The Eclipse MINI Model 98900 is a portable non-invasive continuous ambulatory ECG patchDue to a manufacturing error the incorrect device serial number was potentially programmed into the holter recorder. The physical serial number label on the device might not match the serial number programmed into the device.
-
Class IIofficial
Spacelabs Healthcare, Inc. recalls Xhibit Telemetry ReceiverTelemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts; necessary clinical intervention could be delayed.
-
Class IIofficial
Spacelabs Healthcare, Ltd. recalls Eclipse PRO REF 98700 The Spacelabs model 98700 is a portable non-invasive Holter recorder intended to recordDue to a manufacturing error the incorrect device serial number was potentially programmed into the holter recorder. The physical serial number label on the device might not match the serial number programmed into the device.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions Spacelabs Healthcare, Inc. or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.