Spacelabs Healthcare, Inc. recalls Xhibit Telemetry Receiver with Software
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts and necessary clinical intervention could be delayed.
- Products
-
- Xhibit Telemetry Receiver with Software
- Company
- Spacelabs Healthcare, Inc.
- Where it was sold
-
US: MN, TX, TN, SC, OK, NM, GA, PA, OH, IA, AL, CO, AR, FL, MO, ME, LA, IL, WA, MA, WI, OR, MS, ID, NY, KS, PR. OUS: Canada, Taiwan, Czech Republic, France, Greece, Italy, Kuwait, Philippines, Poland.
States: Alabama, Arkansas, Colorado, Florida, Georgia, Iowa, Idaho, Illinois, Kansas, Louisiana, Massachusetts, Maine, Minnesota, Missouri, Mississippi, New Mexico, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, South Carolina, Tennessee, Texas, Washington, Wisconsin
- Barcodes
10841522107177- Units affected
- 434
- About this recall
- Xhibit Telemetry Receiver with Software, Model 96280. Intended to provide monitoring system with patient data.