SunMed Holdings, LLC recalls
SunMed Holdings, LLC is named on 4 recalls in Recall Relay's records since November 2023, from FDA (15 recalled products: FDA lists each product as its own recall); the most recent was December 15, 2025.
All 15 recalled products, newest first.
-
Class Iofficial
SunMed Holdings, LLC recalls Brand Name: Broselow Product Name: Broselow Pediatric Emergency Rainbow Tape Model/Catalog Number: 7700REA SofProduct contains potentially harmful dosing errors for Vecuronium, Flumazenil, and Ketamine.
-
Class Iofficial
SunMed Holdings, LLC recalls Brand Name: Broselow Product Name: BROSELOW" FILLED BROSELOW ORGANIZER Model/Catalog Number: 7730IALS SoftwareProduct contains potentially harmful dosing errors for Vecuronium, Flumazenil, and Ketamine.
-
Class Iofficial
SunMed Holdings, LLC recalls Brand Name: Broselow Product Name: Broselow Domestic Complete ALS Organizer Model/Catalog Number: 7730ALS SoftProduct contains potentially harmful dosing errors for Vecuronium, Flumazenil, and Ketamine.
-
Class Iofficial
SunMed Holdings, LLC recalls Brand Name: Broselow Product Name: Broselow Pediatric Emergency Rainbow Tape Model/Catalog Number: 7700RE SoftProduct contains potentially harmful dosing errors for Vecuronium, Flumazenil, and Ketamine.
-
Class Iofficial
SunMed Holdings, LLC recalls Broselow Pediatric Emergency Rainbow Tape (distribution by AirLife); REF 7700REA AirLife brandThe impacted tape was manufactured with incorrect information on the tape. Incorrect values are printed in the Red zone, Orange zone, and Grey zone. Using the incorrect values in the Red zone could lead to shocking a patient with an excessive dose of joules, causing significant harm including burns, heart damage, and cardiac arrest. Incorrect values in the Orange zone could lead to overdosing the patient with sodium bicarbonate and cause metabolic alkalosis, electrolyte imbalances, tissue damage, and worsen respiratory status. Incorrect information in the Grey may lead to underdosing the patient with sodium bicarbonate and may cause reduced myocardial contractility, decreased response to vasopressors, and increased risk of dysrhythmia.
-
Class Iofficial
SunMed Holdings, LLC recalls Adult Manual Resuscitator with Medium Adult MaskAffected lots were manufactured with B/V Filter incorrectly attached to the wrong port (patient port instead of the exhalation port). If not noticed prior to patient use, there would be interruption or delay in patient resuscitation, which may lead to life threatening consequences, including hypoxia, hypercapnia, organ failure, and death.
-
Class IIofficial
SunMed Holdings, LLC recalls VentlabA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls Curaplex manual resuscitators with integrated manometerA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls VentlabA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls VentlabA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls VentlabA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls VentlabA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls VentlabA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls VentlabA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls Medline manual resuscitators with integrated manometerA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions SunMed Holdings, LLC or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.
Get recall alerts