SunMed Holdings, LLC recalls Curaplex manual resuscitators with integrated manometer
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
- Products
-
- Curaplex manual resuscitators with integrated manometer
- Company
- SunMed Holdings, LLC
- Where it was sold
-
US Nationwide distribution including in the states of AR, AZ, CA¿CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, NC, ND, NH, NJ, NM, NV, NY, OH, SC, TN, TX, UT, VA, & WV.
Nationwide
- Barcodes
0081007163312600812277037050008122770370740081227703708100812277037067- Units affected
- N/A
- About this recall
- Curaplex manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Curaplex Adult Premium BVM with Bag Reservoir, Manometer, Filter, Peep Valve & Mask, REF # 2442-BVMPAD; 2. Curaplex Child Premium BVM with Bag Reservoir, Manometer, 25cm Pop-off, Filter & Mask, REF # 2442-BVMPCH; 3. Curaplex Child Premium BVM with Bag Reservoir, Manometer, 25cm Pop-off, Filter, Infant and Neonate Mask, REF # 2442-BVMCIN; 4. Curaplex Infant Premium BVM with Bag Reservoir, Manometer, 25cm Pop-off, Filter & 2 Masks, REF # 2442-BVMPIN; 5. Curaplex Small Adult Pediatric Premium BVM with Bag Reservoir, Manometer, 25cm Pop-off, Filter, PEEP Valve and 2 Masks, REF # 2442-BVMPSA.