Sysmex America, Inc. recalls
Sysmex America, Inc. is named on 2 recalls in Recall Relay's records since November 2023, from FDA; the most recent was December 15, 2025.
All 2, newest first.
-
Class IIofficial
Sysmex America, Inc. recalls TS-10/TS-10H Tube SorterUnder specific conditions, the lift's electromagnetic holding force may disengage unexpectedly, allowing unintended downward movement of the lift.
-
Class IIofficial
Sysmex America, Inc. recalls Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6.Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.