Sysmex America, Inc. recalls Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6.
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.
- Products
-
- Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6.
- Company
- Sysmex America, Inc.
- Where it was sold
-
US Nationwide.
Nationwide
- Barcodes
04987562501908- Units affected
- 32 units
- About this recall
- Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Product usage for automated pipetting, diluting, and specimen processing workstations for flow cytometric analysis.