Vyaire Medical recalls
Vyaire Medical is named on 4 recalls in Recall Relay's records since November 2023, from FDA (14 recalled products: FDA lists each product as its own recall); the most recent was April 10, 2025.
All 14 recalled products, newest first.
-
Class Iofficial
Vyaire Medical recalls Brand Name: AirLife Product Name: AirLife Infant Heated Wire Circuit Model/Catalog Number: AH265 Software VersAdapters may disconnect during setup or while in use, potentially interrupting ventilation. Any disconnection can result in severe consequences for neonates, including hypoxia, hypercapnia, and organ failure.
-
Class Iofficial
Vyaire Medical recalls Brand Name: AirLife Product Name: AirLife Infant Heated Wire Circuit Model/Catalog Number: AH165 Software VersAdapters may disconnect during setup or while in use, potentially interrupting ventilation. Any disconnection can result in severe consequences for neonates, including hypoxia, hypercapnia, and organ failure.
-
Class Iofficial
Vyaire Medical, Inc. recalls Vyaire Medical GmbHPotential of the nozzle separating during patient use.
-
Class Iofficial
Vyaire Medical recalls AirLife Adult Manual ResuscitatorRisk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
-
Class Iofficial
Vyaire Medical recalls AirLife Adult Manual ResuscitatorRisk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
-
Class Iofficial
Vyaire Medical recalls AirLife Adult Manual ResuscitatorRisk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
-
Class Iofficial
Vyaire Medical recalls AirLife Adult Manual ResuscitatorRisk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
-
Class Iofficial
Vyaire Medical recalls AirLife Adult Manual ResuscitatorRisk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
-
Class Iofficial
Vyaire Medical recalls AirLife Adult Manual ResuscitatorRisk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
-
Class Iofficial
Vyaire Medical recalls AirLife Adult Manual ResuscitatorRisk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
-
Class Iofficial
Vyaire Medical recalls AirLife Adult Manual ResuscitatorRisk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
-
Class Iofficial
Vyaire Medical recalls AirLife Adult Manual ResuscitatorRisk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
-
Class IIofficial
Vyaire Medical recalls vyaire Flexible Patient CircuitTwo patient circuits, 29028-003 and 29028-004, for the 3100 High Frequency Oscillatory ventilator that are only approved for use outside the U.S. were distributed in the U.S.
-
Class IIofficial
Vyaire Medical recalls vyaire Flexible Patient CircuitTwo patient circuits, 29028-003 and 29028-004, for the 3100 High Frequency Oscillatory ventilator that are only approved for use outside the U.S. were distributed in the U.S.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.